The XlinkX/PD Detect node for cleavable crosslinkers detects spectra containing potential crosslinks on the basis of the characteristic peak doublets, with a constant delta mass resulting from the two fragmentation sites in the crosslink. The following table describes the parameters for this node.
- Set the Acquisition Strategy parameter by selecting the acquisition mode used to acquire the raw data file.
- For the Crosslink Modification parameter, select the appropriate crosslink modification.
If no modification is available, define the crosslink modification in the Chemical Modifications view:
- Click the Crosslink tab in the Edit Chemical Modification dialog box to define the settings for the crosslink modification.
Parameter | Description |
|---|---|
Acquisition Strategy | Specifies the acquisition mode used to acquire the data in the raw data file from the following options:
|
Crosslink Modification | Specifies the crosslink modification used for linking the proteins. If no modification is available, define the crosslink modification in the Chemical Modifications view, including the settings of the crosslink modification in the Modification Extended Properties dialog box. (See Adding crosslinkers.) |
Minimum S/N (Input Data) | Specifies the signal-to-noise spectrum threshold of the MS1 precursor. Default: 1.5 |
Enable protein N-terminus linkage | Specifies the crosslink modification used for linking the proteins. Specifies if a crosslink attachment to the n-terminus of the protein is allowed. Default: false |
SequenceTag Min Length | The minimum length of a sequence tag to consider. Range: 1–100; default: 3 |
SequenceTag Max Length | The maximum length of a sequence tag to consider. Range: 1–100; default: 9 |
SequenceTag Score Filter | Filters out bad sequence tags based on the score when the Noncleavable_open method is used. Default: false |
Minimum S/N (Sequence Tags) | The minimum signal-to-noise accepted for detecting sequence tags when the Noncleavable_open method is used. Default: 10 |
Filter immonium ions | Removes immonium ions prior to sequence tag detection when the Noncleavable_open method is used. Default: true |
Filter duplicate ions | Removes indistinguishable peaks prior to sequence tag detection when the Noncleavable_open method is used. Default: true |