Follow these steps
Press the SERVICE button two times to set the needle in the second service position over the wash port.
Turn off the autosampler with its main power switch.
Remove insulation cover.
Remove the capillary from the insulation cover mounting bracket.
Squeeze the insulation cover mounting bracket and push it upward to remove it.
NOTICE
The torque screwdriver with needle seat bit can get damaged if not used correctly.
Insert the needle seat bit into the torque screwdriver.
Position the torque screwdriver with needle seat bit at the needle seat.
Turn the tool counterclockwise to remove the needle seat.
Unpack the replacement needle seat.
Insert the needle seat and turn it clockwise with your fingers until it is hand-tight.
Use the torque screwdriver with needle seat bit and turn it clockwise to tighten the needle seat until the torque limitation has been reached, which is indicated by a clicking sound.
NOTICE
Needle seat can get damaged when using other tools
The torque screwdriver with needle seat bit is intended for tightening the needle seat. With the help of this tool the needle seat is tightened correctly and will not be overtightened. Do not use any other tool.
Depending on the issue, continue with steps in section Replacing Needle Unit or Completing the Needle Seat and Needle Unit Replacement.