Installation Requirements - Installation Requirements - TSQ Series II Preinstallation Guide HTML - TSQ Quantis Plus MS - TSQ Fortis Plus MS - TSQ Altis MS - TSQ Quantis MS - TSQ Altis Plus MS - TSQ Fortis MS - TSQ Series II Preinstallation Guide HTML - Installation Requirements - TSQ Quantis Plus MS - TSQ Fortis Plus MS - TSQ Altis MS - TSQ Altis MS - TSQ Quantis MS - TSQ Quantis MS - TSQ Altis Plus MS - TSQ Fortis MS - topic

TSQ Series II Preinstallation Guide HTML

Product Tree
Spectrometry > Mass Spectrometry > LC-MS Mass Spectrometers > TSQ Series > TSQ Quantis Plus MS
Spectrometry > Mass Spectrometry > LC-MS Mass Spectrometers > TSQ Series > TSQ Fortis Plus MS
Spectrometry > Mass Spectrometry > IC-MS Mass Spectrometers > TSQ Altis MS
Spectrometry > Mass Spectrometry > LC-MS Mass Spectrometers > TSQ Series > TSQ Altis MS
Spectrometry > Mass Spectrometry > IC-MS Mass Spectrometers > TSQ Quantis MS
Spectrometry > Mass Spectrometry > LC-MS Mass Spectrometers > TSQ Series > TSQ Quantis MS
Spectrometry > Mass Spectrometry > LC-MS Mass Spectrometers > TSQ Series > TSQ Altis Plus MS
Spectrometry > Mass Spectrometry > LC-MS Mass Spectrometers > TSQ Series > TSQ Fortis MS
Document Type
Preinstallation Guide
Software version
4.0
  • TSQ Series II
  • Preface
  • Access Documentation
  • Special Notices, Symbols, and Cautions
  • Vacuum Forepump Types
  • Contact Us
  • Instrument Shipment
  • Inspect Packages for Damage
  • Report Damaged Packages
  • Damage Claims
  • Laboratory Requirements
  • Electrical Requirements
  • Environmental Requirements
  • Gas Supply Requirements
  • System Waste and Exhaust Requirements
  • Installation Requirements
  • Shipping Containers
  • Electrical Ratings and Required Outlets
  • Space and Load Requirements for the Workbenches
  • Requirements for Workbench Clearances
  • Recommended LC/MS layout
  • Customer Supplied Hardware
  • Solvent Requirements and Recommendations
  • Mobile Phase Requirements
  • LC Additive Restrictions
  • Recommended Solvents
  • Line Power Management
  • Power Monitoring Devices
  • Quality of Power
  • Power Conditioning Devices
  • Uninterruptible Power Supply
  • Antivirus Software Management
  • Request the Instrument Installation
  • Instrument Installation and Familiarization
  • Copyright and Trademarks

Before the Thermo Fisher Scientific service engineer installs the MS, read and understand these specific requirements and recommendations.

  • Shipping Containers
  • Electrical Ratings and Required Outlets
  • Requirements for Workbench Clearances
  • Recommended LC/MS layout
  • Customer Supplied Hardware
  • Solvent Requirements and Recommendations
  • Space and Load Requirements for the Workbenches
Back to home page