The Tune application includes built-in instructions for preparing the MS system for the automated System Check and System Calibration processes.
The following topics in this section provide supplemental information for the built-in instructions.
Follow the topics in the order listed.
Procedure
- (Optional) Access system preparation instructions from calibration pane
- Set up syringe pump with the calibration solution
- Position the H-ESI probe or calibration probe for the calibration process
- Apply the favorites settings for the calibration process
- Start the flow of the calibration solution to the ion source
- Check the full scan for missing or interfering peaks
- Evaluate the spray stability