The History pane includes a History preview area and a History Logs area. All changes made to the instrument state that originate from the Tune application are logged in this pane. A change record is automatically created each time you modify a parameter in the Ion Source pane or the Define Scan pane. The record is created after you apply and save the changes. The change record is displayed as a sub-item under the date when it was created and has a maximum limit of 100 change records.
NOTE
A change record is inactive if the source type of the change record differs from the current source type.
Procedure
- Select the History pane.
- Do any of the following:
- - To preview a change record’s parameters, single-click the record in the History area. Parameters shown in bold indicate values that differ from the current settings.
- - To display a change record’s parameters in the Ion Source or Define Scan pane, double-click the record or right-click it and select Load.
- - To submit a change record’s parameters to the MS, double-click the record or right-click it and select Apply. Applying the change creates a new change record.
TIP
Select the arrow on the right side of the window to show or hide the History pane.