The following table below describes the parameters for the Neutral Loss Scan type.

Parameter

Description

Scan Type

The neutral loss scan involves two linked stages of mass analysis, using Q1 and Q3 scanned in parallel over mass ranges of equal width but offset by a specific mass value.

Select Neutral Loss Scan from the drop-down menu.

Scan Range (m/z)

Specify the scan m/z range.

Range: 2–2,010

Neutral Loss (m/z)

Specify the neutral loss.

Range: 2–2,010

Q1 Resolution (FWHM)

Select the Q1 Resolution (FWHM) options from the following options:

Range: 0.2, 0.4, 0.7, 1.2, and 2

Q3 Resolution (FWHM)

Select the Q3 Resolution (FWHM) options from the following options:

Range: 0.2, 0.4, 0.7, 1.2, and 2

Scan Rate (Da/sec)

Select a scan rate from the following options:

Range: 250, 500, and 1,000 (default)

Use Collision Energy Ramp (CER)

Select to turn the feature on or off.

Default: Deselected

CER Start (V)

(Available when the Use Collision Energy Ramp option is selected.)

Specify the Collision Energy Ramp start voltage.

Range: 0–90

CER End (V)

(Available when the Use Collision Energy Ramp option is selected.)

Specify the Collision Energy Ramp end voltage.

Range: 0–90

Collision Energy (V)

Specify the collision energy.

Range: 0–90

CID Gas (mTorr)

Select a CID gas value from the following options:

Range: 0, 0.5, 1, 1.5, 2, 2.5, 3, 3.5, and 4

Source Fragmentation

Specify the collision energy (in volts) for ion source fragmentation.

Range: 0–140