The following table below describes the parameters for the Neutral Loss Scan type.
Parameter | Description |
|---|---|
Scan Type | The neutral loss scan involves two linked stages of mass analysis, using Q1 and Q3 scanned in parallel over mass ranges of equal width but offset by a specific mass value. Select Neutral Loss Scan from the drop-down menu. |
Scan Range (m/z) | Specify the scan m/z range. Range: 2–2,010 |
Neutral Loss (m/z) | Specify the neutral loss. Range: 2–2,010 |
Q1 Resolution (FWHM) | Select the Q1 Resolution (FWHM) options from the following options: Range: 0.2, 0.4, 0.7, 1.2, and 2 |
Q3 Resolution (FWHM) | Select the Q3 Resolution (FWHM) options from the following options: Range: 0.2, 0.4, 0.7, 1.2, and 2 |
Scan Rate (Da/sec) | Select a scan rate from the following options: Range: 250, 500, and 1,000 (default) |
Use Collision Energy Ramp (CER) | Select to turn the feature on or off. Default: Deselected |
CER Start (V) | (Available when the Use Collision Energy Ramp option is selected.) Specify the Collision Energy Ramp start voltage. Range: 0–90 |
CER End (V) | (Available when the Use Collision Energy Ramp option is selected.) Specify the Collision Energy Ramp end voltage. Range: 0–90 |
Collision Energy (V) | Specify the collision energy. Range: 0–90 |
CID Gas (mTorr) | Select a CID gas value from the following options: Range: 0, 0.5, 1, 1.5, 2, 2.5, 3, 3.5, and 4 |
Source Fragmentation | Specify the collision energy (in volts) for ion source fragmentation. Range: 0–140 |