The following table describes the parameters for the SIM Scan type (Q1 and Q3).

Parameter

Description

Scan Type

Selected ion monitoring (SIM) targets specific ions or sets of ions, making it effective for detecting low concentrations of known compounds in complex mixtures.

Select SIM Scan (Q1) or SIM Scan (Q3) from the drop-down menu.

Q1 Resolution (FWHM)

Select the Q1 Resolution (FWHM) options from the following options:

Range: 0.2, 0.4, 0.7, 1.2, and 2

Q3 Resolution (FWHM)

Select the Q3 Resolution (FWHM) options from the following options:

Range: 0.2, 0.4, 0.7, 1.2, and 2

Scan Option

Select the scan type for SIM (Scan Rate or Dwell Time).

Scan Rate (Da/sec)

(Available when the Scan Option is set to Scan Rate.)

Select a scan rate from the following options:

Range: 250 (default), 500, and 1,000.

NOTE: When operating in SIM Scan mode using Q3 and a FWHM of 2, additional scan speeds of 5,000 and 20,000 are available.

Scan Width (m/z)

(Available when the Scan Option is set to Scan Rate.)

Specify the scan width.

Range: 0.5–50

Dwell Time (ms)

(Available when the Scan Option is set to Dwell Time.)

Specify the dwell time in milliseconds.

Range: 0.5–1,000

CID Gas (mTorr)

Select a CID gas value from the following options:

Range: 0, 0.5, 1, 1.5, 2, 2.5, 3, 3.5, and 4

Use Calibrated RF Lens

If the checkbox is selected, the value is taken from the calibration lookup table.

Source Fragmentation

Specify the collision energy (in volts) for ion source fragmentation.

Range: 0–140