The following table describes the buttons and icons located at the top of the Tune window.

Button

Description

On Mode

Indicates that the MS is on and ready for operation. All voltages, sheath gas, auxiliary gas, and sweep gas are active. The vaporizer temperature also heats up, if applicable.

NOTE: The power mode options are available in both the Tune application and data system software. When a power mode is active, the center of the selected icon changes from white to green.

Standby Mode

Indicates that the MS is in standby. In standby, the electron multiplier, conversion dynodes, 8 kV power to the source housing, and ion optic RF voltages are off. The sheath, auxiliary, and sweep gases remain on at a low flow rate (2 arbitrary units) to keep a positive pressure in the ion source to prevent laboratory air from entering the MS.

Standby mode is the normal condition to leave the MS in when it is not in use for short periods of time.

Off Mode

Indicates that the MS is off. When off, all high voltages and the ion gauge are turned off. The sheath, auxiliary, and sweep gases remain on and return to their default settings (2 arbitrary units).

When you remove the source housing or the spray insert, the MS automatically switches to Off mode.

Ion Polarity Mode

Indicates the ion polarity mode (positive or negative). Use the dropdown menu to change the selection.

Data Type

Indicates the data type (profile or centroid). Use the dropdown menu to change the selection.

Scan Averaging

Indicates whether the scan averaging feature is off or on. If scan averaging is turned on, select the dropdown menu to enter a scan averaging value.

Range: 2–1,000; default: 100

NOTE: The Tune application displays the averaged spectrum in a red color. The spectrum header displays the number of averaged scans in the scan filter (for example, AV:2).

Valve Position and Selection

Indicates the current divert/inject valve and its position (1-2 or 1-6). Select the Valve icon to change the valve position, and use the dropdown menu to choose Valve A or Valve B option, as applicable.

Syringe Pump On/Off

Indicates whether the syringe pump is off or on. If the syringe pump is turned on, select the dropdown menu to enter the pump flow rate and syringe volume.

  • Flow rate (µL/min): Varies upon the syringe size
  • Volume: 50, 100, 250, 500, 1,000, 5,000, and 10,000


Holding the Prime button primes the syringe at 100 μL/min.

Data Acquisition Controls

Provides controls for the manual acquisition of sample data. For more details, see Use the Tune application to acquire data.

Browse

Specify a raw file data path for the manual acquisition.

The default path is: C:\ProgramData\Thermo Scientific\Instruments\Data

View

The View button can be linked to a user-specified application, such as the Thermo Scientific FreeStyle™ application, to display the chromatogram and mass spectrum from the raw data file.

To specify the default application, right-click the raw file and select either FreeStyle or Choose Another App.