Procedure

  1. Turn off the SRS Control unit.
  2. Disconnect the analytical (and guard) column(s) from the injection valve and the Dionex SC-CERS 500. Refer to the specific analytical column Installation Instructions and Troubleshooting Guide for column cleanup procedures.
  3. (AutoSuppression External Water Mode) Turn off the external water and disconnect the external water line from the Dionex SC-CERS 500 REGEN IN port.
  4. Disconnect the liquid line from the ELUENT OUT port of the Analyte Converter to the cell at the cell fitting and Reconnect it to the REGEN IN port of the Eluent Suppressor.
  5. Connect a temporary line from the priming block or the low-pressure tee on the isocratic or gradient pump to a container with a solution of freshly prepared 10% 1.0 M methanesulfonic acid (MSA)/90% acetonitrile or methanol. MSA/acetonitrile solutions are not stable during long term storage so this cleanup solution must be made immediately before each suppressor cleanup. Alternatively, the cleaning solution can be proportioned from 1 bottle containing 1.0 M MSA and another bottle containing 100% acetonitrile.
  6. Pump this solution through the Dionex SC-CERS 500 (4 mm) at 1-2 mL/min for 30 minutes. For 2 mm systems, pump this solution through the Dionex SC-CERS 500 (2 mm) at 0.25-0.50 mL/min for 30 minutes.
  7. NOTE

    Bypassing internal pump manifolds when temporarily pumping high concentration cleaning solutions significantly reduces the time required to reequilibrate the system to low concentration eluents.

  8. Flush the Dionex SC-CERS 500 components with deionized water for 10 minutes.
  9. Perform the regeneration procedures described in Regeneration.
  10. Turn on the SRS/ERS Control unit for the AutoSuppression Recycle or External Water Modes of operation. Ensure that the SRS/ERS Control unit is off for the Chemical Suppression Mode of operation.
  11. Flush the Dionex SC-CERS 500 with eluent for 10 minutes.
  12. Reinstall the analytical (and guard) column(s). Begin pumping eluent through the system at the flow rate required for your analysis and equilibrate the system.