NOTE

A Trap Column Suppressor Cleanup Kit (P/N 059659) is recommended for all of the following cleanup procedures.

NOTE

The suppressor voltage is a good indicator of the resistance across the suppressor. Higher resistance may indicate contamination of the suppressor. For more information regarding monitoring the voltage, see Document No 031814-02, “Removal of Iron Contamination from Electrolytic Suppressors.”

Procedure

  1. Turn off the SRS/ERS Control unit.
  2. Disconnect the analytical (and guard) column(s) from the injection valve and the Dionex SC-CERS 500. Refer to the specific analytical column Installation Instructions and Troubleshooting Guide for column cleanup procedures.
  3. (AutoSuppression External Water Mode), turn off the external water and disconnect the external water line from the Dionex SC-CERS 500 REGEN IN port.
  4. Disconnect the liquid line from the ELUENT OUT port of the Analyte Converter to the cell at the cell fitting and reconnect it to the REGEN IN port of the Eluent Suppressor.
  5. Connect a temporary line from the priming block or the low-pressure tee on the isocratic or gradient pump to a container with a solution of 0.2 M oxalic acid. Pump this solution through the Dionex SC-CERS 500 (4 mm) at 1-2 mL/min for 30 minutes. For 2 mm systems pump this solution through the Dionex SC-CERS 500 (2 mm) at 0.25-0.50 mL/min for 30 minutes.
  6. NOTE

    Bypassing internal pump manifolds when temporarily pumping high concentration cleaning solutions significantly reduces the time required to reequilibrate the system to low concentration eluents.

  7. Flush the Dionex SC-CERS 500 components with deionized water for 10 minutes.
  8. Perform the regeneration procedure described in Regeneration.
  9. Turn on the SRS/ERS Control unit for the AutoSuppression Recycle or External Water Modes of operation.
  10. Flush the Dionex SC-CERS 500 components with eluent for 10 minutes.
  11. Reinstall the analytical (and guard) column(s). Begin pumping eluent through the system at the flow rate required for your analysis and equilibrate the system.