If there is leakage from the bottom or side seam of the Dionex SC-CERS 500, check the system back pressure.

Procedure

  1. If the system back pressure is greater than 100 psi, the leaks are caused by excessive back pressure downstream from the Dionex SC-CERS 500. Find and eliminate the source of the pressure.
  2. The Dionex SC-CERS 500 will usually recover from momentary overpressure conditions if allowed to stand approximately 20 minutes with the membranes fully hydrated (see Installation and start up).

If the Dionex SC-CERS 500 continues to leak when operated within the proper back pressure range, it must be replaced. Backpressure greater than 125 psi after the suppressor can cause irreversible damage.