If there is leakage from the bottom or side seam of the Dionex SC-CERS 500, check the system back pressure.
Procedure
- If the system back pressure is greater than 100 psi, the leaks are caused by excessive back pressure downstream from the Dionex SC-CERS 500. Find and eliminate the source of the pressure.
- The Dionex SC-CERS 500 will usually recover from momentary overpressure conditions if allowed to stand approximately 20 minutes with the membranes fully hydrated (see Installation and start up).
If the Dionex SC-CERS 500 continues to leak when operated within the proper back pressure range, it must be replaced. Backpressure greater than 125 psi after the suppressor can cause irreversible damage.