Thermo Fisher Scientific recommends a 1M MSA treatment for the Analyte Converter component. Wear appropriate safety gear during this operation.

Procedure

  1. Disconnect the lines to the device labeled Analyte Converter and perform the following treatment to the Analyte Converter device.
  2. Push 3 mL of 1 M MSA using a syringe through ELUENT IN port and divert a line out of ELUENT OUT port to waste.
  3. Push 5 mL of 1 M MSA using a syringe through REGEN IN port and divert a line out of REGEN OUT port to waste.
  4. Allow the Analyte Converter unit to equilibrate for 20 minutes.
  5. Repeat steps 2 and 3 with DI water and reinstall the unit on the system.