Thermo Fisher Scientific recommends a 1M MSA treatment for the Analyte Converter component. Wear appropriate safety gear during this operation.
Procedure
- Disconnect the lines to the device labeled Analyte Converter and perform the following treatment to the Analyte Converter device.
- Push 3 mL of 1 M MSA using a syringe through ELUENT IN port and divert a line out of ELUENT OUT port to waste.
- Push 5 mL of 1 M MSA using a syringe through REGEN IN port and divert a line out of REGEN OUT port to waste.
- Allow the Analyte Converter unit to equilibrate for 20 minutes.
- Repeat steps 2 and 3 with DI water and reinstall the unit on the system.