Procedure
- Check that the Dionex SC-CERS is operating at the correct current (see Recommended current settings), and that there are no suppressor alarms.
- Check for eluent flow out of the ELUENT OUT port of the Eluent Suppressor component of the Dionex SC-CERS 500.
- If there is no flow out of the ELUENT OUT port, make sure that eluent is entering the Dionex SC-CERS 500 at the ELUENT IN port. If there is no flow at this point, trace the eluent flow path backward through the system to find and remove the blockage.
- If there is flow into the Dionex SC-CERS 500 but not out, and there are no visible leaks from the side seam of the suppressor, a break in the membrane is probably allowing eluent to leak into the regenerant chambers. If this is the case, then the Dionex SC-CERS 500 must be replaced. The Dionex SC-CERS 500 is sealed during manufacture. Attempting to open it will destroy it.
- If there is flow into and out of the Eluent Suppressor, then check the Analyte Converter for eluent flow out by following the above steps.
CAUTION
Do not attempt to disassemble the Dionex SC-CERS 500.
- If there is flow from the ELUENT OUT ports of both suppressors, but no eluent suppression, the membranes may have been contaminated. Try to restore system performance by cleaning the membranes (see Cleanup).
- Remake the eluent to be sure that the concentration is correct. Be sure that chemicals of the required purity were used to make the eluent (see Chemical purity requirements).
- If the eluent concentration is high, the Dionex SC-CERS 500 may not be set up to suppress this high concentration, resulting in high background conductivity. See Recommended current settings to determine the correct amount of current that should be applied to a Dionex SC-CERS 500.
- Increase the power setting on the suppressor. Raising the current above 300 mA for a 4mm ES suppressor or above 100 mA for a 2mm ES suppressor will shorten the suppressor lifetime.