Procedure
- Verify that the Dionex SC-CERS 500 is connected to the power supply and the current is set properly (see Recommended current settings).
The Electrolytically Regenerated Suppressor does not support electrolysis: There is no conductive liquid in the eluent and/or the regenerant chamber(s). Causes of this malfunction may include the following:
- Lack of eluent and regenerant flow.
- Plugged or leaking tubing.
- Plugged or leaking or fittings.
- Lack of pneumatic pressure on the eluent or water bottles in the AutoSuppression External Water Mode.
- Empty eluent or water bottles.
- Analytical pump is not operating.
- Eluent is non-ionic liquid.