The Dionex RFC-10 suppressor controller is required to operate the Dionex SC-CERS 500 in IC systems without Chromeleon and a Dionex ICS system with suppressor power supply. Refer to the Dionex RFC-10 operator's manual for details on its installation and operation.
When the Dionex RFC-10 module status says “not enabled” even through the pump is on, perform the following steps to solve the problem.
Procedure
- Check the TTL-2 OUT connector for a loose wire.
- Check that the pump TTL2 output is set to 0 Flow in the Pump Options screen.