ED50A, CD25A, and IC25A modules in Dionex DX-600 systems equipped with Chromeleon 6.2 SP2 or later chromatography workstation provides full operation control to the Dionex SC-CERS 500. When using these modules, the suppressor type and current setting will be shown on the module front panel.
The "Open circuits are detected from the suppressor" error message can be caused by incorrect identification of the suppressor being used or the suppressor is damaged.
Procedure
- Make sure that the SRS cable is connected properly to the Dionex SC-CERS 500 suppressor, the CERS suppressor type is selected, and the eluent flow is on.
If the problem persists, it is possible that the Dionex SC-CERS 500 suppressor is damaged or fouled. See Cleanup. Check and ensure it is identified as a CERS suppressor.