Long-term storage (more than 5 days) - Long-term storage (more than 5 days) - SC-CERS 500 Suppressor User Guide - SC-CERS 500 - Long-term storage (more than 5 days) - SC-CERS 500 - topic
SC-CERS 500 Suppressor User Guide
- Product Tree
- Chromatography > Ion Chromatography > Suppressors > SC-CERS 500
- Document Type
- User Guide
- Attach the storage line (P/N 035936) that is shipped with the Dionex SC-CERS 500 to the REGEN OUT port.
- Flush the Dionex SC-CERS 500 with deionized water for 10 minutes.
- Plug all Dionex SC-CERS 500 eluent ports. Connect the filled storage line (P/N 035936) to the REGEN IN port.
- To resume operation, perform a quick start before connecting the suppressor to the system. Allow the system to equilibrate before starting analysis.