For optimal suppressor performance, it is important to use the recommended suppressor currents for the target applications. The Program Wizard of the Dionex Chromatography Management System (see Control Programmed Program Wizard) guides you in setting the recommended current for the Dionex SC-CERS 500.

The Wizard supplies a recommended current based on the eluent concentration and flow rate settings (See Recommended current settings to determine the correct amount of current that should be applied to a Dionex SC-CERS 500.). The Dionex Chromatography Management System automatically enters the recommended current into the Program and applies that current to the Dionex SC-CERS 500.

When setting the current to the Dionex SC-CERS 500 using the Chromeleon Wizard, choose CERS as the suppressor type.

NOTE

In Chromeleon, the suppressor current could also be entered manually by editing the program file. For calculating the current manually, refer to Dionex SC-CERS 500 current settings.