AC - Analyte Converter Quick Start - AC - Analyte Converter Quick Start - SC-CERS 500 Suppressor User Guide - SC-CERS 500 - AC - Analyte Converter Quick Start - SC-CERS 500 - topic
SC-CERS 500 Suppressor User Guide
- Product Tree
- Chromatography > Ion Chromatography > Suppressors > SC-CERS 500
- Document Type
- User Guide
- Disconnect the lines to the component labeled Analyte Converter.
- Attach a temporary waste line from the ELUENT IN port to a waste container.
- Use a disposable plastic syringe push approximately 3 mL of deionized water through the ELUENT OUT port.
- Attach a temporary waste line from the REGEN OUT port to a waste container.
- Push approximately 5 mL of deionized water through the REGEN IN port.
- Allow the AC component to sit for approximately 20 minutes to fully hydrate the suppressor membranes and screens.