Procedure

  1. Disconnect the lines to the component labeled Analyte Converter.
  2. Attach a temporary waste line from the ELUENT IN port to a waste container.
  3. Use a disposable plastic syringe push approximately 3 mL of deionized water through the ELUENT OUT port.
  4. Attach a temporary waste line from the REGEN OUT port to a waste container.
  5. Push approximately 5 mL of deionized water through the REGEN IN port.
  6. Allow the AC component to sit for approximately 20 minutes to fully hydrate the suppressor membranes and screens.