Procedure
- Disconnect the lines to the component labeled Analyte Converter.
- Attach a temporary waste line from the ELUENT IN port to a waste container.
- Use a disposable plastic syringe push approximately 3 mL of 200 mN MSA through the ELUENT OUT port.
- Attach a temporary waste line from the REGEN OUT port to a waste container.
- Push approximately 5 mL of 200 mN MSA through the REGEN IN port.
- Allow the AC component to sit for approximately 20 minutes to fully hydrate the suppressor membranes and screens.
- Displace the 200 mN MSA by pushing 3 mL of DI water through the ELUENT OUT port and 5 mL of DI water through the REGEN IN port.
CAUTION
Do not cap the waste reservoir. The very small amount of hydrogen gas generated by the Dionex SC-CERS 500 is not dangerous unless the gas is trapped in a closed container and allowed to accumulate. The Gas Separator Waste Tube must be open to the atmosphere and not in a confined space to operate properly.
- After regeneration is complete, do the following:
- Mount the suppressors and plug in the ES power cable. Plumb the Dionex SC-CERS 500 components according to the mode of operation.
- Turn on the power to the Dionex SC-CERS 500 and establish eluent flow through the Dionex SC-CERS 500. Be sure to turn on the SRS or ERS Control power and the pump flow at the same time.
CAUTION
Always turn the pump and the SRS/ERS Control on and off at the same time.
CAUTION
Eluent flow through the Dionex SC-CERS 500 is required for proper operation. However, without current, the membranes and screens in the Dionex SC-CERS 500 will become expended of regeneration ions by the flowing eluent, resulting in small analyte peak areas. If this should occur, perform the procedure outlined in Decreased sensitivity.
- Start operation. Allow the system to equilibrate before beginning analysis.