Procedure

  1. Disconnect the lines to the component labeled Analyte Converter.
  2. Attach a temporary waste line from the ELUENT IN port to a waste container.
  3. Use a disposable plastic syringe push approximately 3 mL of 200 mN MSA through the ELUENT OUT port.
  4. Attach a temporary waste line from the REGEN OUT port to a waste container.
  5. Push approximately 5 mL of 200 mN MSA through the REGEN IN port.
  6. Allow the AC component to sit for approximately 20 minutes to fully hydrate the suppressor membranes and screens.
  7. Displace the 200 mN MSA by pushing 3 mL of DI water through the ELUENT OUT port and 5 mL of DI water through the REGEN IN port.
  8. CAUTION

    Do not cap the waste reservoir. The very small amount of hydrogen gas generated by the Dionex SC-CERS 500 is not dangerous unless the gas is trapped in a closed container and allowed to accumulate. The Gas Separator Waste Tube must be open to the atmosphere and not in a confined space to operate properly.

  9. After regeneration is complete, do the following:
  10. Mount the suppressors and plug in the ES power cable. Plumb the Dionex SC-CERS 500 components according to the mode of operation.
  11. Turn on the power to the Dionex SC-CERS 500 and establish eluent flow through the Dionex SC-CERS 500. Be sure to turn on the SRS or ERS Control power and the pump flow at the same time.
  12. CAUTION

    Always turn the pump and the SRS/ERS Control on and off at the same time.

  13. CAUTION

    Eluent flow through the Dionex SC-CERS 500 is required for proper operation. However, without current, the membranes and screens in the Dionex SC-CERS 500 will become expended of regeneration ions by the flowing eluent, resulting in small analyte peak areas. If this should occur, perform the procedure outlined in Decreased sensitivity.

  14. Start operation. Allow the system to equilibrate before beginning analysis.