Procedure

  1. Disconnect the lines to the component labeled Eluent Suppressor.
  2. Attach a temporary waste line from the ELUENT IN port to a waste container.
  3. Use a disposable plastic syringe push approximately 3 ml of 200 mN NaOH through the ELUENT OUT port.
  4. Attach a temporary waste line from the REGEN OUT port to a waste container.
  5. Push approximately 5 ml of 200 mN NaOH through the REGEN IN port.
  6. Allow the ES component to sit for approximately 20 minutes to fully hydrate the suppressor membranes and screens.
  7. Displace the 200 mN NaOH by pushing 3 mL of DI water through the ELUENT OUT port and 5 mL of DI water through the REGEN IN port.