For applications requiring high temperatures (40°C to 60°C), the Dionex SC-CERS 500 must be installed outside the chromatography oven; add a length of tubing (up to 20”, 50 cm) between the column outlet and the suppressor inlet to allow time for the eluent to cool to room temperature if operating the column above 40°C.

Use a minimal length of tubing from the column outlet to the ELUENT IN port of the Dionex SC-CERS 500 and from the REGEN OUT port to the conductivity cell. The tubing should have the ends square cut so no dead volume is introduced.

The line from the cell to the REGEN IN port should be 1/8” i.d. to minimize backpressure to the Dionex SC-CERS 500.

When using a temperature controlled chromatography module, a conductivity cell with a shield (P/N 044132) should be placed inside the oven.

CAUTION

The Conductivity Cell with DS3 (P/N 044130) will not compensate for temperatures above 45°C. Placing the DS3 Cell outside the oven will lead to poor sensitivity since the Dionex SC-CERS 500 and DS3 are at a lower temperature than the oven.