The Dionex RFC-10 and Dionex RFC-30 units can be used to supply current to the Dionex SC-CERS 500 in 1 mA increments. Refer to the Reagent-Free Controller Operator’s Manual (Document # 031880) for detailed instructions.
Select the CERS_2MM or CERS_4MM for Dionex SC-CERS operation and use the current settings for CERS operation.
NOTE
If using DX 1-03, it is recommended to use a Dionex RFC-10 or Dionex RFC-30 power supply to provide more accurate current to the Dionex SC-CERS 500
The 2-position rotary switch located on the back of the RFC-30 allows the selection of either 110 V or 220 V input line voltage. It can be adjusted with a large flat-bladed screwdriver or a small coin. The unit is not frequency sensitive and will work equally well on 50 or 60 Hz without adjustment. Each position accepts a wide range of input line voltages.
- Use the 110 V position for all input line voltages between 85 VAC and 135 VAC.
- Use the 220 V position for all input line voltages between 175 VAC and 265 VAC.
The control and display panel on the Dionex RFC-30 is located on the front of that module’s cabinet. LED indicators inform the operator of the system’s operational status.
The Dionex RFC-30 has its own VAC Power On/Off switch to toggle the VAC line current to the Dionex RFC-30 on and off. This is the only way to turn the current output to the Dionex SC-CERS 500 on and off on the Dionex RFC-30. See the following figure.
Two line input fuses are located in the fuse drawer on the socket of the VAC input connector on the back panel of the RFC-30.
To replace fuses, first remove the VAC line cord, and then remove the fuse drawer by squeezing the retaining clips located at the right and left of the drawer. Pull the drawer outward to replace both fuses. Spare fuses are found in the shipping kit. The fuses are 5 x 20 mm, 0.315A/250V, FAST IEC127 (P/N 954747).
NOTE
It is recommended to use a Dionex RFC-10 or Dionex RFC-30 power supply to provide more accurate current to the Dionex SC-CERS 500.
CAUTION
As a general operating precaution, never apply current to the Dionex SC-CERS 500 without eluent or water regenerant flowing through the Dionex SC-CERS 500 at the same time. Always apply current whenever eluent is running through the suppressor.