SRS/ERS control for these instruments is accessible from the detector front panel or PeakNet/Chromeleon software. For information on operation from PeakNet/Chromeleon, please see the Help file in the Run or Method files.
NOTE
If using the ED40, first select Conductivity Mode from the Main Screen.
NOTE
If using the Dionex DX-600, DX-500, or Dionex DX-320, it is recommended to use a Dionex RFC-10 or Dionex RFC-30 power supply to provide more accurate current to the Dionex SC-CERS 500.
To operate the SRS/ERS from the front panel of the detector, follow the procedure below.
Procedure
- Select Main Screen or Detail Screen to ensure the detector is in one of the main conductivity screens.
- Select Menu until you see this as an option. Press the appropriate number, then press Enter.
- Make sure you are in 1, the Local mode. If you are in the Remote mode, use the arrow keys to select Local, then press Enter.
- In the Main or Detail screens, use the arrow keys to navigate to the SRS field.
- Set the current output level. Increase or decrease the levels by using the select keys, labeled 50, 100, 300, and 500 mA.
- Press Enter.
Task | Procedure |
|---|---|
SRS/ERS Control Connections for the CD20/ED40 | Procedure
|
SRS/ERS Line Voltage for the CD20/ED40 | These detectors have automatic switching power supplies to adjust to the line voltage. No user settings are required. |
SRS/ERS Power Control for the CD20/ED40 | The CD20 and ED40 are controlled from the instrument front panel on Dionex DX-500 instruments. Procedure
|
CAUTION
As a general operating precaution, never apply current to the Dionex SC-CERS 500 without eluent or water regenerant flowing through the Dionex SC-CERS 500 at the same time. Always apply current whenever eluent is running through the suppressor.
NOTE
The Dionex SC-CERS 500 is not compatible with the Dionex DX-120 power supply. A Dionex RFC-10 or RFC-30 power supply is recommended for use with Dionex DX-120 instruments.
See the following figure and table for details on CDM-3/PED-2 control (Dionex DX-300 instruments).
|
|
|---|---|
Detector Power Switch | Located on the rear panel. Powers SRS/ERS control when turned on. |
Cell OFF/ON | Located on the front panel. Turns current flow on or off to both the cell and the Dionex SC-CERS 500. |
SRS Control Current Selector | Located on the top panel of the detector. |
Function | Description |
|---|---|
SRS/ERS Control connections | Use the SRS/ERS Control extension cable (P/N 045343) to connect the SRS/ERS Control output connector on the back of the detector chassis, to the Dionex SC-CERS 500 current input connector. |
SRS/ERS Line Voltage | Receives its power directly from the CDM-3 or PED-2 detector. If the detector's voltage control (CORCOM) is properly set (see the appropriate CDM-3 or PED-2 Operator's Manual), the control portion of the detector will receive the correct power without further adjustment. |
SRS/ERS Power Control for The CDM-3/PED-2 (Dionex DX-300 Instruments) | The control and display panels on the CDM-3 and PED-2 detectors for the Dionex SC-CERS 500 are located under the top cover, visible through an opening in the cover. LED indicators inform the operator of the system’s operational status. The CDM-3 and the PED-2 detector VAC Power On/Off switch, located on the rear panel of the instrument, toggles the VAC Line voltage to the CDM-3 or PED-2 detector and the internal SRS/ERS Control on and off. SRS/ERS Control is therefore powered whenever the detector is powered. To interrupt the current to the Dionex SC-CERS 500 while the detector is On, use the detector cell On/Off control. With the detector cell On, the detector conductivity cell is On and the detector SRS/ERS Control provides current to the SC-CERS 500. With the detector cell Off, the detector conductivity cell is Off and the detector SRS/ERS Control output current to the Dionex SC-CERS 500 is Off although the detector SRS/ERS control still receives power. |
NOTE
If using the Dionex DX-500, it is recommended to use a Dionex RFC-10 or RFC-30 power supply to provide more accurate content to the Dionex SC-CERS 500.
CAUTION
As a general operating precaution, never apply current to the Dionex SC-CERS 500 without eluent or water regenerant flowing through the Dionex SC-CERS 500 at the same time. Always apply current whenever eluent is running through the suppressor.