This section describes the available detectors. For information on selecting current settings, see Dionex SC-CERS 500 current settings.

Electrolytically Regenerated Suppressor Control consists of the electronic components necessary to control and provide conditioned power to the Dionex SC-CERS 500 Suppressors.

CAUTION

Always turn the pump and ERS Control on and off at the same time. Eluent flow through the Dionex SC-CERS 500 is required for proper operation. However, without current, the membranes and screens in the Dionex SC-CERS 500 will become expended by the flowing eluent resulting in anomalous analyte peak areas and baseline drift.

If this should occur, perform the procedure outlined in Nonlinear response or loss of sensitivity.

The Dionex SC-CERS 500 can also be controlled using the Dionex RFC-10 Power Supply Module which has the capabilities of supplying the optimum current to the Dionex SC-CERS 500. See Using a Dionex RFC-10 or Dionex RFC-30 power supply to power the Dionex SC-CERS 500 for information on using the Dionex RFC-10 to control the Dionex SC-CERS 500.