The AutoSuppression Recycle Mode uses the suppressed conductivity cell effluent as the source of water for the regenerant chamber water. This is the recommended method of operation for the Dionex SC-CERS 500.
The advantage of this mode of operation is its simplicity and ease of use. It reliably provides salt conversion for most suppressed conductivity applications using MSA eluents. As the eluent passes through the Dionex SC-CERS 500, it is substantially neutralized to form a low background. After passing through the conductivity cell, the effluent is redirected to the regenerant inlet on the Dionex SC-CERS 500 Eluent Suppressor, thus supplying it with a source of water containing low levels of MSA and minute amounts of diluted analyte. See the following figure.
The amount of water flowing through the regenerant chambers is therefore equal to the eluent flow rate. Because of this limitation, the AutoSuppression Recycle Mode of operation cannot be used with eluents containing organic solvents. See Operation for more information.