The Dionex SC-CERS 500 mode of operation depends mainly on the eluent composition.
For example, eluents containing organic solvents are not compatible with the AutoSuppression Recycle Mode; the AutoSuppression External Water Mode should be used instead.
The Eluent Suppressor component of the Dionex SC-CERS 500 uses dilute MSA and electrolysis as the regenerant to achieve substantial eluent suppression.
The most common mode of operation is the AutoSuppression Recycle Mode. In this mode of operation, substantially suppressed eluent flows from the eluent outlet of the suppressor into the conductivity cell and is then recycled through the Dionex SC-CERS 500 regenerant chambers. This eliminates the need for an external source of water but restricts the regenerant flow rate to the eluent flow rate.
The Analyte Converter component uses the MSA effluent from the Eluent Suppressor as a regenerant to achieve conversion of the analytes to MSA. The AutoSuppression External Water Mode incorporates an external source of deionized water flowing through the regenerant chambers. This requires the installation of a water delivery system such as a pressurized bottle system or peristaltic pump to provide an external source of water. With this configuration, the regenerant flow rate is not restricted to the eluent flow rate.
Both the AutoSuppression Recycle Mode and the AutoSuppression External Water Mode require application of electric current to the Eluent Suppressor component of the Dionex SC-CERS 500.
To improve performance and increase suppressor lifetime, the optimum current must be applied. The optimum current setting produces just enough hydroxide ions to displace the eluent counter ions, thus neutralizing the eluent. When using solvents in the eluent, the optimum current setting is critical for prevention of oxidation of the solvent.
See Recommended current settings.
Eluent Composition | AutoSuppression Recycle | AutoSuppression External Water |
|---|---|---|
MSA Eluents | Yes | Yes |
Sulfuric Acid and Other Eluents | No | No |
MSA Eluent Containing Organic Solvents | No | Yes |
Eluents Containing Ion Pair Reagents | No | No |