NOTE

Make sure that Xcalibur 4.7 is connected to the Ardia Platform before starting the setup.

For more information on the Ardia Repository parameters, see Ardia Repository parameters in the Ardia Platform Help portal.

Prerequisites

  1. You must register Proteome Discoverer prior to registering for automated processing. See Register on the Ardia Server.
  2. Xcalibur 4.7 must be connected to the Ardia Platform before starting.

Procedure

  1. Open the Xcalibur data system.
  2. Select Ardia Platform in the Roadmap View.
  3. Select Open in the Xcalibur Registration card.
  4. The Ardia Platform Registration dialog box opens.
  5. Enter the Platform URL and Registration Code. To find the URL and code, do the following:
  6. Open a browser and navigate to the Ardia Platform Home page on the Ardia Advanced Tower Server.
  7. Select the Client Registration and Management card.
  8. The Actions view opens.
  9. Select + Add Registration Code in the upper right corner of the Actions view.
  10. The You have added the registration code successfully dialog box opens.
  11. Select Copy to copy the Platform URL.
  12. Return to the Xcalibur data system and paste the URL in the Platform URL box.
  13. Return to the Ardia Platform browser and select Copy to copy the Registration Code.
  14. Return to the Xcalibur data system and paste the code in the Registration Code box.
  15. Select Continue.
  16. The registration is successful.
  17. Select Close.
  18. Sign in to the Ardia Platform.
  19. Select My Account in the upper right corner and select Sign In.
  20. In the Xcalibur data system, select Sequence Setup View in the View toolbar.
  21. Two new options are now available in the Data Repository area.
  22. In the Data Repository area, select the Ardia Platform option.
  23. In the Data Repository area, select the Ardia Platform processing with Proteome Discoverer software option.
  24. The following three new columns appear: Processing Method, Quantification Method, and Results Path.
  25. From the Processing Method dropdown list and select the processing method.
  26. The list contains all of the analysis templates that you saved for automated processing. If the computer is on a shared server, you will also see templates created on other computers.
  27. From the Quantification Method dropdown list, select the quantification method.
  28. Double-click the Path column.
  29. The Sequence Selector dialog box opens.
  30. Select a folder or create a new folder where your raw data will be stored.
  31. To create a new folder, select New folder.
  32. A new row appears.
  33. Enter a name for the folder and double-click the row.
  34. The folder opens.
  35. Enter a sequence name in the Sequence Name box.
  36. Select Save.
  37. Double-click the Results Path column.
  38. The Data Explorer dialog box opens.
  39. Navigate to the results folder where your results will be stored.
  40. Select Open.
  41. Select Run Sample or Run Sequence .
  42. To view the acquisition status, select the Ardia Platform connection status in the upper right corner.
  43. Select Open in the Acquisition and Upload Status card.
  44. The Acquisition and Upload Status page appears and displays the status of the current sequence file.
  45. After the Sequence Status and Upload Completed columns both display completed, then you can go back to Proteome Discoverer to view the results.
  46. In the Proteome Discoverer application, open the Job Queue Administration > Show Job Queue.
  47. The runs that completed are listed in the Job Queue.
  48. Under Ardia Server Management, select Automated Processing.
  49. In the Processing Statistics area, you will see that the Last 24 hours column has changed.
  50. To open your result file in Proteome Discoverer, you can open a result or add the result file from the Ardia Advanced Tower Server to the study.
  51. In Proteome Discoverer, on the Analysis Results page of the study, select Add Files.
  52. Select Add Remote Results.
  53. The Pick Documents dialog box opens.
  54. Navigate to the result file.
  55. Select Open.
  56. The results appear on the Analysis Results. A green checkmark appears in the Remote Status column.