Install the source diagnostic cartridge

Use the source diagnostic cartridge to test the camera views in the Tune application, perform installation diagnostics, and check the OptiSpray system tightness.

TIP

For a video demonstration of this process, see the following video.

WARNING - MAGNETIC FIELD HAZARD

The OptiSpray ion source generates strong magnetic fields that can interfere with your device.

If you have a pacemaker or implanted cardioverter defibrillator (ICD), maintain a minimum distance of 6 inches (15 centimeters) between the OptiSpray ion source and your implanted device.

Prerequisites

  • OptiSpray ion source and sweep cone are installed.

  • Shipping jig has been removed from the OptiSpray ion source.

  • Vanquish Neo UHPLC pump is turned off and fluid is not flowing to the OptiSpray ion source.

Procedure

  1. Lift the handle on the OptiSpray ion source.
    Figure 1. OptiSpray handle (raised position)
    OptiSpray handle (raised position)
  2. Retract the protective cover on top of the source diagnostic cartridge to expose the connection ports.
    Figure 2. Retract the protective cover on the cartridge
  3. With the connection ports facing up, insert the source diagnostic cartridge into the cartridge slot.
  4. Firmly pull the handle down.

    The cartridge LED turns white, the cartridge initiates a homing routine, and the cartridge information is displayed in the Tune application.

What to do next

Continue to Test the camera views in the Tune application.