Continue to Install the infusion cartridge.
Remove the source diagnostic cartridge
After testing the OptiSpray system tightness, remove the source diagnostic cartridge.
TIP
For a video demonstration of this process, see the following video.
Procedure
-
From the Tune ribbon, select the Cartridge type drop-down menu.
Figure 1. Select the source diagnostic cartridge in the Tune application
The cartridge details pane opens.
-
Select Remove Cartridge.
Figure 2. Remove Cartridge button in the Tune application
-
After the pressure on the column is below the specified limits, select OK on the dialog box that opens.
If the cartridge is in the active position, it moves to the inactive position.
CAUTION
The cartridge pressure must be below 10 bar (150 PSI) before you remove the cartridge. Removing the cartridge at high pressure might cause damage.
Figure 3. Cartridge removal dialog box -
After the cartridge status indicator on the source housing turns off, place the clamping handle in the unlocked (up) position.
NOTE
If the clamping handle does not lift up with gentle pressure, press the handle down firmly, then select Retry on the cartridge removal dialog box.
- Remove the source diagnostic cartridge from the OptiSpray ion source. Slide the protective cover back over the top of the cartridge to cover the connection ports, then store the cartridge in a cool, dry location.