Test the camera views in the Tune application

Verify that both camera views are in focus and that the camera lighting modes are functioning properly.

NOTE

The source diagnostic cartridge is designed for running automated diagnostics and does not have an emitter.

TIP

For a video demonstration of this process, see the following video.

Prerequisites

  • OptiSpray ion source, sweep cone, and source diagnostic cartridge are installed.

Procedure

  1. In the Tune application, ensure that the mass spectrometer is set to Standby mode.
  2. From the right side of the Tune window, select Camera.
  3. Verify that both camera lighting modes are functioning:
    1. From the header bar of the Camera pane, select Lighting to change the lighting mode between white (cartridge visualization) and green (spray visualization).
      Figure 1. OptiSpray camera lighting modes

      Cartridge visualization (white lighting)Spray visualization (green lighting)
  4. Verify that both camera views are in focus:
    1. From the Tune ribbon, select the cartridge status drop-down menu.
    2. Select Active.
      The cartridge moves in front of the MS system. The ion transfer tube should be visible from the Camera pane.

      NOTE

      If the ion transfer tube is not visible or in focus, open the Ion Source page and select the Emitter Optimization tab. Set the optimization type to Manual, then use the single-arrow or double-arrow icons to incrementally move the cartridge.

      Figure 2. Manual optimization with XYZ movement type
      Manual optimization with XYZ movement type

What to do next

Continue to Run installation diagnostics.