Install the OptiSpray ion source

The OptiSpray ion source is designed for quick installation without the use of additional tools.

Install the OptiSpray sweep cone on the mass spectrometer, then attach the OptiSpray ion source and connect the required cables.

WARNING - MAGNETIC FIELD HAZARD

The OptiSpray ion source generates strong magnetic fields that can interfere with your device.

If you have a pacemaker or implanted cardioverter defibrillator (ICD), maintain a minimum distance of 6 inches (15 centimeters) between the OptiSpray ion source and your implanted device.

Required components

  • OptiSpray ion source

  • OptiSpray sweep cone (P/N B51000127)

  • Power adapter (P/N B51003488)

  • Ethernet cable (P/N B51008613)

  • USB-C cables (P/N B51003217)

TIP

For a video demonstration of this process, see the following video.

Prerequisites

The mass spectrometer is set to Standby or Off mode in the Tune application.

The API source interface has cooled to room temperature.

Install the OptiSpray sweep cone on the mass spectrometer

CAUTION - HOT SURFACE

Heated components of the API source interface, including the OptiSpray sweep cone and the FAIMS interface, can reach 350° C (662° F).

Set the MS system to Off mode in the Tune application, then allow components to cool to 55° C (131° F) before handling. Cooling takes approximately 20 minutes.

Procedure

  1. If the standard sweep cone is installed, remove it.
  2. Push the OptiSpray sweep cone onto the API source interface.
    Figure 1. Align and install the OptiSpray sweep cone onto the API source interface
  3. Ensure that the bore hole is centered around the ion transfer tube. Adjust the alignment by gently pressing one side, if necessary.
  4. Insert and tighten the screws (clockwise) on both sides of the sweep cone to ensure a proper fit and to protect the emitter from damage.
    Figure 2. OptiSpray sweep cone installed on the MS system
    OptiSpray sweep cone installed on the MS system

Attach the OptiSpray ion source to the mass spectrometer

Procedure

  1. Move the locking levers on the back of the OptiSpray ion source to the unlocked (down) position.
    Figure 3. OptiSpray locking levers and guide pin holes.
    12
    1. OptiSpray locking levers

    2. OptiSpray guide pin holes

  2. Align the OptiSpray guide pin holes with the guide pins on the mass spectrometer.
    Figure 4. Align the OptiSpray ion source with the mass spectrometer

    Align the OptiSpray ion source with the mass spectrometer
  3. Carefully slide the OptiSpray ion source onto the mass spectrometer.
  4. Move the OptiSpray locking levers to the locked (up) position.
  5. Connect the following cables:
    • Ethernet cable to instrument hub

    • Power cable to the 24 V power supply

    • USB-C cables to the instrument computer

    CAUTION

    Securely place the 24 V power supply on a flat surface, or place it in the syringe pump compartment of the Orbitrap Astral, Orbitrap Exploris, or Orbitrap Excedion mass spectrometer.

    Do not allow the power supply to hang from its cable. Doing so can damage the OptiSpray ion source power port.

    NOTE

    Extension cords and longer USB-A to USB-C cables are available, if needed. See the Replaceable parts section of the OptiSpray user guide for ordering information.

  6. Connect a fluidic line from the LC system to the HPLC port.
    • Finger-tighten the Thermo Scientific™ Viper™ fitting until it is snug.

    • Do not tighten more than one-quarter turn after you feel resistance.

    CAUTION

    Over-tightening can damage the Viper fitting.

    CAUTION

    When you use the OptiSpray ion source with the Vanquish Neo UHPLC system, use the Vanquish Neo torque tool (P/N 6250.2110) to tighten the Viper fitting. This ensures optimal performance at pressures up to 1,500 bar.

    Figure 5. Connection ports on the OptiSpray ion source
    Connection ports on the OptiSpray ion source
  7. Turn on the OptiSpray ion source.
  8. Verify that the ion source LED fully illuminates white, which indicates initialization is complete.

NOTE

If the system cannot establish communication within 7 minutes, the connection attempt times out. The LED remains in the breathing white state, but Tune continues to show the system as disconnected.

If this occurs, see OptiSpray cartridge cannot be removed in the OptiSpray user guide for troubleshooting procedures.

Remove the shipping jig

Procedure

  1. Locate the zip tie on the clamping handle.
  2. Press and hold the release tab on the zip tie, then gently slide the tie out to remove it.

    CAUTION

    Avoid pulling the tie forcefully to prevent it from snapping back when released.

  3. Lift the clamping handle on the OptiSpray ion source to unlock it.
  4. Pull the shipping jig out, keeping it perpendicular to the OptiSpray ion source.

    WARNING

    To prevent damage to the OptiSpray ion source, do not flow solvent from the HPLC system without an OptiSpray cartridge installed.

What to do next

Continue to Install the source diagnostic cartridge.