Connect the Vanquish Neo system to the EASY-Spray column - Connect the Vanquish Neo system to the EASY-Spray column - EASY-Spray Series Ion Source User Guide - EASY-Spray Source - Connect the Vanquish Neo system to the EASY-Spray column - EASY-Spray Source - topic

EASY-Spray Series Ion Source User Guide

Product Tree
Spectrometry > Mass Spectrometry > LC-MS Mass Spectrometers > Accessories > EASY-Spray Source
Document Type
User Guide
  • EASY-Spray Series Ion Source User Guide
  • Preface
  • Supported instruments
  • Intended use
  • Safety precautions
  • Special notices, symbols, and cautions
  • Access product documentation
  • Contact us
  • Introduction
  • Nanospray ionization
  • Ion source components
  • Source housing
  • Temperature controller
  • Controller power supply
  • Specifications
  • Install the ion source
  • Connect power to the ion source
  • Connect power using an EASY-nLC 1200 system
  • Connect power using the controller power supply
  • Attach the ion source
  • Attach the Tee holder
  • Adjust the emitter tip position
  • Remove the ion source
  • Install the USB camera
  • Adjust the video picture
  • Install and connect the EASY-Spray column
  • Tools and supplies
  • NanoViper fitting guidelines
  • Install the EASY-Spray column
  • Connect the Vanquish Neo system to the EASY-Spray column
  • One-column Vanquish Neo plumbing configuration
  • Two-column Vanquish Neo plumbing configuration
  • Connect the EASY-nLC 1200 system to the EASY-Spray column
  • One-column EASY-nLC 1200 plumbing configuration
  • Two-column EASY-nLC 1200 plumbing configuration
  • Two-column EASY-nLC 1200 software configuration
  • Connect the RSLCnano system to the EASY-Spray column
  • One-column RSLCnano plumbing configuration
  • Two-column RSLCnano plumbing configuration
  • Set the column temperature
  • Configure the MS for NSI mode
  • Configure the EASY-Spray NG source for NSI mode
  • Configure the EASY-Spray source for NSI mode
  • Maintenance
  • Guidelines
  • Tools and supplies
  • Clean the EASY-Spray column emitter tip
  • Replace the temperature controller PCB
  • Troubleshooting
  • Parts and accessories
  • Installation kit contents
  • Legal notices
  • Copyright and trademarks
  • Regulatory compliance
Jump to main content
EASY-Spray Series Ion Source User Guide
Logo
  1. Home
  2. Install and connect the EASY-Spray column
  3. Connect the Vanquish Neo system to the EASY-Spray column
  • EASY-Spray Series Ion Source User Guide
  • Preface
  • Introduction
  • Install the ion source
  • Remove the ion source
  • Install the USB camera
  • Install and connect the EASY-Spray column
    • Tools and supplies
    • NanoViper fitting guidelines
    • Install the EASY-Spray column
    • Connect the Vanquish Neo system to the EASY-Spray column
      • One-column Vanquish Neo plumbing configuration
      • Two-column Vanquish Neo plumbing configuration
    • Connect the EASY-nLC 1200 system to the EASY-Spray column
    • Connect the RSLCnano system to the EASY-Spray column
    • Set the column temperature
  • Configure the MS for NSI mode
  • Maintenance
  • Troubleshooting
  • Parts and accessories
  • Legal notices

Connect the Vanquish Neo system to the EASY-Spray column

Connect the Vanquish Neo UHPLC system plumbing to the EASY-Spray column by using either a one-column or two-column plumbing configuration.

  • One-column Vanquish Neo plumbing configuration

  • Two-column Vanquish Neo plumbing configuration

  • One-column Vanquish Neo plumbing configuration
  • Two-column Vanquish Neo plumbing configuration