Connect power using the controller power supply

Use the controller power supply supplied with the EASY-Spray Series ion source to power the ion source when using a Thermo Scientific™ Vanquish™ Neo UHPLC system or any nanoLC system other than the EASY-nLC™ 1200.

Procedure

  1. Turn the ion source temperature control knob to OFF.
  2. If necessary, connect the detachable input plug for your country or region and power adapter cable to the power supply.

    CAUTION

    To avoid electric shock, use only the detachable input plug supplied with the power supply that is appropriate for your country or region.

  3. Align the red dot on the connector with the red dot on the right side of the P-BUS socket, and then connect the power adapter cable to the bottom of the ion source.

    IMPORTANT

    Before connecting the plug, visually confirm that the red dots on the plug and socket are aligned. Do not connect the plug unless you can confirm the alignment.

  4. Plug the power supply into an electrical wall outlet.