Do not overtighten the fittings. Overtightening can permanently damage the sealing surfaces. Use the Viper torque screwdriver to achieve a leak-free connection at pressures up to 1,500 bar.
Figure 1. Trapping column installed in the Vanquish Neo autosampler Required tools and materials
EASY-Spray column
Precolumn/trap column
Viper torque screwdriver (P/N 6250.2110)
Viper union (P/N 6040.2304)
Prerequisites
Wear appropriate personal protective equipment (PPE), including a lab coat, safety glasses, and chemical-resistant gloves.
Procedure
If the Vanquish Neo system is not already configured for a two-column setup, run the A06 - Change Workflow / Fluidics script and select Trap-and-Elute.
The script guides you through the fluidic changes. Follow the prompts to install the second precolumn/trap column between port 4 and port 6 of the Vanquish Neo autosampler valve.
Install the Viper union on the inlet of the EASY-Spray column.
Connect the transfer line from the Vanquish Neo autosampler valve to the Viper union installed on the EASY-Spray column.
Tighten the fitting by using the Viper torque screwdriver.
NOTE
For information about operating the Vanquish Neo system and using Vanquish Neo scripts to configure the direct injection workflow, analytical column type, and analytical column specifications, refer to the Vanquish Neo Operating Manual and the Vanquish Neo User Guide.