Nanospray ionization
Nanoelectrospray ionization (NSI or nanoESI) is a form of electrospray ionization (ESI) that operates at low flow rates, typically 10 to 1,000 nL/min. Compared to conventional ESI, NSI produces smaller droplets that improve desolvation efficiency and increase analyte concentration at the emitter tip. As a result, NSI provides higher sensitivity while typically eliminating the need for drying gas or additional thermal heating.
Conventional ESI typically operates at flow rates between 1 μL/min and 1 mL/min. At these higher flow rates, the larger volume of solvent often requires drying gas, thermal heating, or both to promote droplet desolvation. In contrast, the lower flow rates used by NSI reduce solvent volume, enabling efficient ionization over a wider range of mobile-phase compositions, including pure water.
In ESI, a high voltage applied to the emitter or solvent creates an electric field at the emitter tip. This field charges the liquid as it exits the emitter, producing an aerosol of charged droplets that ultimately generates gas-phase ions for mass spectrometric analysis. The lower flow rates used by NSI produce smaller droplets, increasing ionization efficiency and improving analytical sensitivity.