Perform check teaching in the User level.

For the PAL terminal button functions, refer to the PAL terminal section.

Steps

  1. On the PAL terminal screen, select Tray Holder 1.
  2. Select tray holder 1
    Select tray holder 1
  3. Select Options and select Check Teaching.
  4. Select check teaching
    Select check teaching
  5. The on-screen graphic displays the teaching points for the 4-position tray holder (there are two teach points). For checking the first teach point, press Check. The robot arm moves to the teach point and checks teaching by lowering the tool to the teach point and verifying the correct position and coordinates along the X, Y, and Z axes. When positioned correctly, the tool aligns with the center of the teach point position.
  6. Check first teach point
    Check first teach point
    4-position tray holder first teach point
    4-position tray holder first teach point
  7. When done, press Next.
  8. Press next
    Press next
  9. Press Check to check the teaching on the second teach point.
  10. Check second teach point
    Check second teach point
    4-position tray holder second teach point
    4-position tray holder second teach point
  11. When the teaching is checked, press Next on the PAL terminal.
  12. Press next