Run an experiment - Run an experiment - AccelerOme™ User Guide - AccelerOme - AccelerOme™ User Guide - Run an experiment - Running experiments on the instrument - AccelerOme - AccelerOme - topic

AccelerOme™ User Guide

Product Tree
Spectrometry > Mass Spectrometry > Sample Preparation > AccelerOme
Chromatography > Sample Preparation > AccelerOme
Document Type
User Guide
  • Preface
  • Access Documentation
  • Special Notices
  • Special Notices, Symbols, and Cautions
  • Contact Us
  • Introduction
  • Product information
  • Type plate
  • Document information
  • Disclaimer and restrictions of use
  • Intended use of the AccelerOme system
  • Limit of Detection (LOD) for UV performance
  • Incorrect usage and exclusion of liability
  • Foreseeable misconduct and misuse
  • Residual risks and dangers
  • AccelerOme system description
  • Design and general functionality
  • Instrument front and back design
  • Power switch
  • Power and ethernet cord connections
  • Instrument front connections
  • Solution compartment
  • Service compartment
  • Front door
  • Instrument interface
  • Instrument interior deck
  • Robot arm
  • Dilutor pump and tool
  • Tool change station
  • Tray slots
  • Thermal mixer
  • Wash station
  • Barcode scanner
  • Dimensions and weight
  • Laboratory requirements
  • Laboratory entrance requirements
  • Space and load requirements
  • Power supply specifications
  • Power line specifications
  • Electrical specifications
  • Laboratory table
  • Emissions
  • Technical data
  • Dilutor
  • Thermal mixer
  • Tray cooler (optional)
  • Environmental conditions, limits for operation, and storage
  • Network requirements
  • General specifications
  • Experiment Designer (desktop software)
  • Optional components
  • Tray cooler module
  • Basic safety information
  • Hazards
  • Regulatory and safety
  • Regulatory compliance
  • FCC compliance statement
  • Notices
  • Prepare the instrument for use
  • Transportation and storage
  • Tools required for installation
  • Unpack the instrument
  • Package contents
  • Installation
  • Preinstall Instructions
  • Preinstallation check
  • Solution preparation
  • Make the electrical connections
  • Connect the waste line to the waste container
  • Load the needle wash solutions
  • Install consumables and system components
  • Install the dilutor syringe
  • Install the dilutor needle
  • Install the barcode scanner
  • Install the tray inserts and solution tray adapters
  • Prepare the samples
  • Sample performance specifications
  • Install the well plates, buffer, SPE cartridge, and reagent plates
  • Power on the system
  • System set up
  • Install the Experiment Designer software
  • Set up the instrument interface
  • Experiment Designer (desktop software)
  • System requirements for Experiment Designer
  • Design a method file
  • Design an LFQ or TMT method
  • Import method
  • Sample information
  • Factor definition
  • Plate setup
  • Statistical power estimate
  • Batch or sequence information
  • Sample randomization
  • File review
  • Experiment Designer parameters
  • Home screen parameters
  • Method file parameters
  • Edit samples parameters
  • Define factors parameters
  • Setup plate parameters
  • Assign batches/sequences parameters
  • Review and send parameters
  • Manage instruments
  • Add an instrument
  • Remove an instrument
  • Monitor an instrument
  • Instrument parameters
  • History
  • Method design history
  • History parameters
  • Settings
  • Default save location
  • Software updates
  • Statistical review
  • File name convention
  • About
  • Settings parameters
  • Running experiments on the instrument
  • Instrument safety features
  • Safety symbols on the instrument
  • Door lock information
  • Instrument interior illumination
  • Acoustic announcement movements
  • Robot arm movement control
  • Hazardous substances information
  • Run an experiment
  • Instrument interface home screen parameters
  • Run an experiment using a method file
  • Run an experiment using quick start
  • Install the IQ kit
  • Instrument interface features
  • Manage users
  • Add a user
  • Password requirements
  • Deactivate a user
  • Reactivate a user
  • Change user profile information
  • Add user parameters
  • Reconstitution for LCMS
  • Change the reconstitution for LCMS
  • Software overview
  • System configuration
  • Service tools
  • Prime or empty the lines
  • Perform a system test
  • Perform a thermomixer test
  • Change the needle of the dilutor tool
  • Access the event log
  • Event log retention policy
  • Set the tray cooler temperature
  • Perform an install qualification test
  • Re-measure qualification samples
  • View, delete or export qualification test history
  • Shut down the instrument
  • Service tools menu items
  • Load the instrument deck for use
  • Consumables and spare parts
  • Spare parts
  • Consumables
  • AccelerOme system consumables
  • AccelerOme system needle wash solutions
  • AccelerOme system reagent kit guides
  • Disposal of consumables
  • Waste fluid disposal
  • Preventative maintenance
  • Required tools and materials
  • Clean the instrument
  • Check and replace the solution tubes
  • Check and replace the waste line
  • Replace the dilutor syringe
  • Replace the needle of the dilutor tool
  • Check teaching and install qualification test
  • Prepare instrument for long-term storage
  • Preventative maintenance schedule
  • Decontamination and disposal
  • Decontaminate the instrument
  • Disposal
  • Repack the instrument for shipment
  • Troubleshooting
  • Troubleshooting robot arm issues
  • Troubleshooting dilutor valve and syringe issues
  • Troubleshooting SPE cartridge availability issue
  • Troubleshooting dilutor tool configuration issue
  • Troubleshooting device access failure
  • Troubleshooting instrument interface access issues
  • Troubleshooting power on self test failure
  • Troubleshooting deck validation and process run failure
  • Troubleshooting needle guide homing issue
  • Troubleshooting instrument fan issues
  • Troubleshooting Experiment Designer and instrument pairing
  • Proteome Discoverer troubleshooting
  • Troubleshooting improper shut down issue
  • PAL subsystem reference
  • PAL terminal
  • General conventions
  • Paths
  • Description of menu items
  • The menu screen
  • Select an item or a command from a list
  • Options menu
  • Select an item from a drop-down list
  • Enter text on the PAL terminal
  • Access levels
  • User level
  • Extended user level
  • Service level
  • Change the access level
  • Example of differences in access levels
  • Status bar symbols
  • Content area symbols
  • Check teaching
  • Check tool calibration
  • Check vial positions
  • Check tool change station teaching
  • Check 4-position tray holder teaching
  • Check needle wash teaching
  • Check solid waste teaching
  • Check thermal mixer teaching
  • Check tray cooler teaching
  • Check tray A teaching
  • Check tray B1 teaching
  • Check tray B2 teaching
  • Check tray D buffers teaching
  • Check tray SPE teaching
  • Teaching
  • Tool calibration
  • Tool change station teaching
  • 4-position tray holder teaching
  • Needle wash teaching
  • Solid waste teaching
  • Thermal mixer teaching
  • Tray cooler teaching
  • Tray A teaching
  • Tray B1 teaching
  • Tray B2 teaching
  • Tray D buffers teaching
  • Tray SPE teaching
  • Robot arm calibration
  • Copy a backup file to a USB drive
  • Install the IQ kit and the solution tray adapters
  • Install the tray inserts and solution tray adapters
  • Install the IQ kit
  • Run the install qualification test
  • AccelerOme Automated Sample Preparation System
  • Copyright
  • Regulatory Compliance
  • FCC Compliance Statement
  • WEEE Directive
  • Notices

There are two ways to run an experiment on the AccelerOme instrument:

  • Run an experiment using a method file
  • Run an experiment using quick start
Back to home page