The technical data of the tray cooler is as follows:

Technical data for the tray cooler module

Specification

Technical data

Part number

Part number of module: C0960-01-00111

Temperature control

4 °C (39.2 °F) to 40 °C (104 °F) in 0.1 °C increments. Peltier element for cooling functionality of the device.

For the AccelerOme system, the supported temperature range for the tray cooler is 6 °C (42.8 °F) to 17 °C (62.6 °F).

Temperature cooling

T∆ = ≥ 14.0 °C within 60 minutes.

T∆ = TAmbient temperature – TTray Cooler

T∆max = 22°C (for example, for an ambient temperature of 32°C (89.6 °F), a minimum temperature of 10 °C (50 °F) can be reached).

The value of TTray Cooler can be checked in the extended user level using the parameter Tray Cooler Temp.